Inst ToolsInst ToolsInst Tools
  • Ask
  • Courses
  • Videos
  • Q & A
    • Interview
      • Instrumentation
      • Electronics
      • Electrical
      • Practical Questions
    • MCQ
      • Instrumentation MCQ
      • Electrical MCQ
      • Electronics MCQ
      • Control Systems MCQ
      • Analog Electronics MCQ
      • Digital Electronics MCQ
      • Power Electronics MCQ
      • Microprocessor MCQ
      • Multiple Choice Questions
  • EE
    • Electronics
      • Electronics Q & A
      • Electronic Basics
      • Electronic Devices & Circuits
      • Electronics Animation
      • Digital Electronics
    • Electrical
      • Electrical Basics
      • Electrical Q & A
      • Power Electronics
      • Electrical Machines
      • Electrical Animation
      • Power Systems
      • Switchgear & Protection
      • Transmission & Distribution
  • Measure
    • Control Valves
    • Calibration
    • Temperature
    • Pressure
    • Flow
    • Level
    • Analyzers
    • Switches
    • Vibration
    • Solenoid Valve
  • Control
    • PLC Tutorials
    • Control Systems
    • Safety Instrumented System (SIS)
    • Communication
    • Fire & Gas System
  • More
    • Design
    • Tools
    • Animation
    • Basics
    • Formulas
    • Standards
    • TextBooks
    • Common
    • Software
    • Excel Tools
    • Erection & Commissioning
    • Process Fundamentals
    • Videos
    • Books
Search
All rights reserved. Reproduction in whole or in part without written permission is prohibited.
Reading: Why it is important to note As-Found and As-Left data during Calibration?
Share
Notification Show More
Font ResizerAa
Inst ToolsInst Tools
Font ResizerAa
  • Courses
  • PLC Tutorials
  • Control Systems
Search
  • Ask
  • Courses
  • Videos
  • Q & A
    • Interview
    • MCQ
  • EE
    • Electronics
    • Electrical
  • Measure
    • Control Valves
    • Calibration
    • Temperature
    • Pressure
    • Flow
    • Level
    • Analyzers
    • Switches
    • Vibration
    • Solenoid Valve
  • Control
    • PLC Tutorials
    • Control Systems
    • Safety Instrumented System (SIS)
    • Communication
    • Fire & Gas System
  • More
    • Design
    • Tools
    • Animation
    • Basics
    • Formulas
    • Standards
    • TextBooks
    • Common
    • Software
    • Excel Tools
    • Erection & Commissioning
    • Process Fundamentals
    • Videos
    • Books
Follow US
All rights reserved. Reproduction in whole or in part without written permission is prohibited.
Inst Tools > Blog > Calibration > Why it is important to note As-Found and As-Left data during Calibration?

Why it is important to note As-Found and As-Left data during Calibration?

Last updated: September 2, 2020 5:31 pm
Editorial Staff
Calibration
No Comments
Share
3 Min Read
SHARE

An instrument technician working for a pharmaceutical processing company is given the task of calibrating a temperature recording device used to display and log the temperature of a critical batch vessel used to grow cultures of bacteria.

Contents
Solution:As-Found and As-Left dataRead Next:

After removing the instrument from the vessel and bringing it to a workbench in the calibration lab, the technician connects it to a calibration standard which has the ability to simulate a wide range of temperatures. This way, he will be able to test how the device responds to different temperatures and make adjustments if necessary.

Before making any adjustments, though, the technician first inputs the full range of temperatures to this instrument to see how it responds in its present condition. Then, the instrument indications are recorded as As-Found data. Only after this step is taken does the technician make corrections to the instrument’s calibration. Then, the instrument is put through one more full-range test and the indications recorded as As-Left data.

Explain why it is important that the technician make note of both “As-Found” and “As-Left” data? Why not just immediately make adjustments as soon as an error is detected? Why record any of this data at all? Try to think of a practical scenario where this might matter.

Solution:

I’ll answer the question with a scenario of my own: suppose it is discovered that some patients suffered complications after taking drugs manufactured by this company and that the particular batch of suspect drugs was processed in this very same vessel about 6 months ago? Now imagine that this temperature recording instrument gets routinely calibrated once a month. See the problem?

Read: As-found and As-left Documents

Share your thoughts with us through comments.

As-Found and As-Left data

The below diagrams shows the example As-Found and As-Left data for a temperature and pressure transmitter that are connected to primary and standby flow computers. (for reference)

Temperature Transmitter As-Found and As-Left Document

As-Found and As-Left Calibration

Credits: Tony R. Kuphaldt

Read Next:

  • What is Calibration?
  • Instrument Zero and Span Calibration
  • How to do Transmitter Calibration?
  • Instruments Calibration Procedure
  • Questions & Answers on Calibration
Don't Miss Our Updates
Be the first to get exclusive content straight to your email.
We promise not to spam you. You can unsubscribe at any time.
Invalid email address
You've successfully subscribed !
Single Position Calibration of Temperature Bath
H & B Gas Analyzer Principle and Calibration Procedure
Calibration : Up-tests and Down-tests
Multiphase Flow Meter Calibration Procedure
Shunt Calibration of a Strain Gauge Load Cell
Share This Article
Facebook Whatsapp Whatsapp LinkedIn Copy Link
Share
Leave a Comment

Leave a Reply Cancel reply

Your email address will not be published. Required fields are marked *

Stay Connected

128.3kFollowersLike
69.1kFollowersFollow
208kSubscribersSubscribe
38kFollowersFollow

Categories

Recent Comments

  • Kamli on Top Free PLC Software
  • Guifty Shimica on Top Non-PLC Certification Courses for Automation Professionals
  • Guifty Shimica on Top Non-PLC Certification Courses for Automation Professionals
  • MIHARITSOA Aina Sitraka on Top Non-PLC Certification Courses for Automation Professionals

Related Articles

RTD Testing Methods

RTD Testing Methods in Calibration Laboratory

Pressure transmitter Calibration Setup

Instruments Calibration Procedures

what is calibration

Important Terms in Calibration

Sample Raw DataSheet for Pressure Calibration

Sample Raw DataSheet for Pressure Calibration

Specific-Conductance-Calibration-Procedure

Specific Conductance Calibration Procedure

How to Calibrate Load Cells

How to Calibrate Load Cells in Weighing Applications?

Temperature Switch Calibration

How to Calibrate Temperature Switch? – Instrument Basics

RTD Calibration

RTD Calibration

More Articles

Polyvinyl Alcohol in Oil Field

Polyvinyl Alcohol in Oil Field

Redundant Power Supply WITH Diode Unit

Redundant Power Supply for Critical Field Panels

Gas Detection System Abbreviations

Gas Detection System Abbreviations

Multistage Flash Drums Erratic Level Readings

Multistage Flash Drums Erratic Level Readings

Current Divider Rule

Current Divider Rule

Mass Spectrometer Questions and Answers

Surface Spectroscopic Techniques

Power Electronics Objective Questions

Quiz on Type E Choppers

Induction Motor Torque

AC Induction Motor Slip

Follow US
All rights reserved. Reproduction in whole or in part without written permission is prohibited.
Welcome Back!

Sign in to your account

Username or Email Address
Password

Lost your password?