Inst ToolsInst ToolsInst Tools
  • Courses
  • Automation
    • PLC
    • Control System
    • Safety System
    • Communication
    • Fire & Gas System
  • Instrumentation
    • Design
    • Pressure
    • Temperature
    • Flow
    • Level
    • Vibration
    • Analyzer
    • Control Valve
    • Switch
    • Calibration
    • Erection & Commissioning
  • Interview
    • Instrumentation
    • Electrical
    • Electronics
    • Practical
  • Q&A
    • Instrumentation
    • Control System
    • Electrical
    • Electronics
    • Analog Electronics
    • Digital Electronics
    • Power Electronics
    • Microprocessor
  • Request
Search
  • Books
  • Software
  • Projects
  • Process
  • Tools
  • Basics
  • Formula
  • Power Plant
  • Root Cause Analysis
  • Electrical Basics
  • Animation
  • Standards
  • 4-20 mA Course
  • Siemens PLC Course
Reading: Chemiluminescence
Share
Notification Show More
Font ResizerAa
Inst ToolsInst Tools
Font ResizerAa
  • Courses
  • Design
  • PLC
  • Interview
  • Control System
Search
  • Courses
  • Automation
    • PLC
    • Control System
    • Safety System
    • Communication
    • Fire & Gas System
  • Instrumentation
    • Design
    • Pressure
    • Temperature
    • Flow
    • Level
    • Vibration
    • Analyzer
    • Control Valve
    • Switch
    • Calibration
    • Erection & Commissioning
  • Interview
    • Instrumentation
    • Electrical
    • Electronics
    • Practical
  • Q&A
    • Instrumentation
    • Control System
    • Electrical
    • Electronics
    • Analog Electronics
    • Digital Electronics
    • Power Electronics
    • Microprocessor
  • Request
Follow US
All rights reserved. Reproduction in whole or in part without written permission is prohibited.
Inst Tools > Blog > Analyzers > Chemiluminescence

Chemiluminescence

Last updated: September 11, 2018 10:39 am
Editorial Staff
Analyzers
No Comments
Share
7 Min Read
SHARE

Recall that an exothermic chemical reaction is one that releases a net sum of energy, as opposed to an endothermic reaction which requires a greater input of energy than it releases. Combustion is a common class of exothermic reactions, with the released energy being very obviously in the forms of heat and light, with heat being the predominant form.

Some exothermic reactions release energy primarily in the form of light rather than heat. The general term for this effect is chemiluminescence. A natural example is the “cold” light emitted by North American species of firefly. In this small insect, a chemical reaction intermittently takes place emitting significant amounts of light but insignificant amounts of heat. An artificial example is the light emitted by a “glow stick” when activated.

The following photographs show such a light source before activation when the reactants are separated (left) and after activation when the internal barrier is broken and the reactants are allowed to mix (right):

Chemiluminescence

Certain industrial compounds engage in chemiluminescent reactions, and this phenomenon may be used to measure the concentration of those compounds. One such compound is nitric oxide (NO), an atmospheric pollutant formed by high-temperature combustion with air as the oxidizer.

A spontaneous chemical reaction between nitric oxide and ozone (an unstable molecule formed of three oxygen atoms: O3) is known to produce chemiluminescence:

NO + O3   →   NO2 + O2 + light

Although this process of generating light is quite inefficient (only a small fraction of the NO2 molecules formed by this reaction will emit light), it is predictable enough to be used as a quantitative measurement method for nitric oxide gas. Ozone gas is very easy to produce on demand, by exposing air or oxygen to a high-voltage electric discharge.

A simplified diagram for a chemiluminescent nitric oxide gas analyzer appears here:

Chemiluminescent Analyzer Principle

As with many optical analyzers, a photomultiplier tube serves as the light-detecting sensor, generating an electrical signal in proportion to the amount of light observed inside the reaction chamber. The higher the concentration of NO molecules in the sample gas stream, the more light will be emitted inside the reaction chamber, resulting in a stronger electrical signal produced by the photomultiplier tube.

Although this instrument readily measures the concentration of nitric oxide (NO), it is insensitive to other oxides of nitrogen (NO2, NO3, etc., collectively referred to as NOx, pronounced “nocks”). Normally, we would consider this selectivity to be a good thing, because it would eliminate interference problems from these other gases. However, as it so happens, these other oxides of nitrogen are every bit as polluting as nitric oxide, and therefore when we measure nitric oxide for pollution monitoring purposes, we usually also wish to measure these other oxides in combination.

In order to use chemiluminescence to measure all oxides of nitrogen, we must chemically convert the other oxides into nitric oxide (NO) before the sample enters the reaction chamber. This is done in a special module of the analyzer called a converter :

Chemiluminescent Analyzer

A three-way solenoid valve is shown in this diagram, providing a means to bypass the converter so the analyzer only measures nitric oxide content in the sample gas. With the solenoid valve passing all the sample through the converter, the analyzer responds to all oxides of nitrogen (NOx) and not just nitric oxide (NO).

One simple way to achieve the NOx → NO chemical conversion is to simply heat the sample gas to a high temperature, around 1300o F. At this temperature, the molecular structure of NO is favored over more complex oxides such as NO2, the result being a release of oxygen from the NO2 and NO3 molecules to become NO molecules. A disadvantage of this technique is that those same high temperatures also have a tendency to convert other compounds of nitrogen such as ammonia (NH3) into nitric oxide, thereby creating an unintended interference species.

An alternative NOx → NO conversion technique is to use a metallic reactant in the converter to remove the extra oxygen atoms from the NO2 molecules. One such metal that works well for this purpose is molybdenum (Mo) heated to the comparatively low temperature of 750 oF, which is too low to convert ammonia into nitric oxide. The reaction of NO2 converting to NO is as follows:

3NO2 +Mo → MoO3 + 3NO

Other oxides (such as NO3) convert in similar fashion, leaving their excess oxygen atoms bound to molybdenum atoms and becoming nitric oxide (NO). The only difference between these reactions and the one shown for NO2 is the proportional (stoichiometric) ratios between molecules. As you can see from the reaction, the molybdenum metal is converted into the compound molybdenum trioxide over time, requiring periodic replacement. The rate at which the molybdenum metal depletes inside the converter depends on the sample flow rate and the concentration of NO2.

As with other optical gas analyzers, pressure control of the gas sample is critically important for good measurement accuracy. If the pressure of the sampled gas inside the chemiluminescence reaction chamber happens to vary, it will affect the amount of light emitted even if the relative concentration of NOx gas remains stable. This is because higher pressures will pack gas molecules closer together, resulting in more reactive molecules inside the chamber for any given percentage or ppm concentration. For this reason you will see analyzers such as this equipped with pressure regulation to ensure the gas pressure inside the measurement chamber remains constant.

Don't Miss Our Updates
Be the first to get exclusive content straight to your email.
We promise not to spam you. You can unsubscribe at any time.
Invalid email address
You've successfully subscribed !

Continue Reading

Dissolved oxygen probe calibration procedure
NDIR Analyzer Filter Cells
Hydrazine Analyzer Working Principle
Single-beam Non-dispersive Analyzer
H & B Gas Analyzer Principle and Calibration Procedure
pH measurement
Share This Article
Facebook Whatsapp Whatsapp LinkedIn Copy Link
Share
Leave a Comment

Leave a Reply Cancel reply

Your email address will not be published. Required fields are marked *

Stay Connected

128.3kFollowersLike
69.1kFollowersFollow
210kSubscribersSubscribe
38kFollowersFollow

Categories

Explore More

Basics of Chromatograph Operation
Ammonia Gas Detector Working Principle and Calibration
Oil in Water Analyzer Working Principle
pH Analyzer Calibration and Troubleshooting
Basics of Total Organic Carbon (TOC) Analyzers
Oxygen Analyzer Working Principle
How to Calibrate Oxygen Analyzer? – O2 Sensor Testing Procedure
Thin-Layer Chromatography : Manual Method

Keep Learning

Total Organic Carbon Analyzers Questions and Answers

Total Organic Carbon Analyzers Questions and Answers

Ambient Air Quality Monitoring System Principle

Ambient Air Quality Monitoring System Principle

Suspended Particulate Matter

Basics of Suspended Particulate Matter (SPM) Analyzers

Dissolved Oxygen Analyzers Interview Questions

Dissolved Oxygen Analyzer Interview Questions & Answers

Calibrate Eddy Current Conductivity Instrument

How to Calibrate the Eddy Current Conductivity Meter?

H2 S and Total Sulfur Using Lead Acetate Principle

H2S Measurement Using Lead Acetate Principle

How to Calibrate pH Electrode?

How to Calibrate pH Electrode?

In-Situ Zirconia Oxygen Analyzer

Oxygen (O2) Measurement in Flue Gas

Learn More

Introduction to SCL Programming in Siemens PLC

Types of Expressions in the SCL Language of the Tia Portal

Vacuum Gauges Working Principle

Vacuum Gauges Working Principle

Tagging Philosophy for Junction Box

Tagging Philosophy for Junction Box, Cable, and Cable Drum

Bipolar Junction Transistors Questions & Answers

Measurement and Instrumentation Objective Questions – Part 1

How to Test a Battery using Multimeter

How to Test a Battery using a Multimeter?

PLC based 4 Way Traffic Light Control System

PLC based 4 Way Traffic Light Control System

Comparison of Control Loops - Open, Closed, Cascade, Feedforward, Ratio

Comparison of Control Loops – Open, Closed, Cascade, Feedforward, Ratio

Top 100 Power Systems Projects

Top 100 Power Systems Projects – Electrical Engineering

Menu

  • About
  • Privacy Policy
  • Copyright

Quick Links

  • Learn PLC
  • Helping Hand
  • Part Time Job

YouTube Subscribe

Follow US
All rights reserved. Reproduction in whole or in part without written permission is prohibited.
Welcome Back!

Sign in to your account

Username or Email Address
Password

Lost your password?