Here we discuss the brief overview of different In-House testing methods carried out for RTD assemblies in the Calibration Laboratory.
The below list shows the different testing methods.
RTD calibration can be done as per IEC 751 / DIN 43760 Class A & B.
Normally test is conducted for two points (i.e. 00 C & 1000 C), 3 point calibration or more can be followed.
It should be more than 100 M ohms.
IR should be more than 2 M ohms. More than 20 M ohms also can be offered.
For the RTD sensor, after cap welding, the N2 leakage test should be carried out no leakage should be observed at 40 kg/cm2.
Response time test as per IEC 751 – (63.2% step change from ambient to 800C).
For RTD without transmitter – 6 to 9 sec up to 6 mm OD of sheath.
For RTD with thermowell – the response time will be 35 to 45 sec.
Autoclave test – RTD kept in steam @ 1250C & at 1.2 kg/cm2 for about 1 hour.
IR should be more than 5 M Ohms.
By using continuity tester/multimeter – This test is used to confirm the element is connected properly to the RTD transducer cable.
Water Immersion test – To check the sheath integrity of thermocouple / MI RTD cable.
Response time test for RTD with RTD connected to pipe on its outside surface to which water is flowing @ 20 ltrs. per second & @ 800C. The acceptance norm is less than 20 seconds.
It is in accordance with IEC 751
source: Prisma Instruments
PLC ladder logic design to control 3 motors with toggle switch and explain the program…
VFD simulator download: Master the online tool from the Yaskawa V1000 & programming software for…
The conveyor sorting machine is widely used in the packing industries using the PLC program…
Learn the example of flip-flop PLC program for lamps application using the ladder logic to…
In this article, you will learn the STAR DELTA programming using PLC controller to start…
Lube oil consoles of rotary equipment packages in industrial process plants are usually equipped with…
View Comments
This is very important my knowledge