Reading: Calibration and Troubleshooting of Oxygen Analyzer
Share
Notification Show More
Font ResizerAa
Inst ToolsInst Tools
Font ResizerAa
  • Courses
  • Design
  • PLC
  • Interview
  • Control System
Search
  • Courses
  • Automation
    • PLC
    • Control System
    • Safety System
    • Communication
    • Fire & Gas System
  • Instrumentation
    • Design
    • Pressure
    • Temperature
    • Flow
    • Level
    • Vibration
    • Analyzer
    • Control Valve
    • Switch
    • Calibration
    • Erection & Commissioning
  • Interview
    • Instrumentation
    • Electrical
    • Electronics
    • Practical
  • Q&A
    • Instrumentation
    • Control System
    • Electrical
    • Electronics
    • Analog Electronics
    • Digital Electronics
    • Power Electronics
    • Microprocessor
Follow US
All rights reserved. Reproduction in whole or in part without written permission is prohibited.
Inst Tools > Blog > Analyzers > Calibration and Troubleshooting of Oxygen Analyzer

Calibration and Troubleshooting of Oxygen Analyzer

Learn how to do Zero and Span Calibration and Troubleshooting of Oxygen Analyzer used in the industries.

Last updated: November 9, 2020 11:40 am
Editorial Staff
Analyzers Calibration
2 Comments
Share
3 Min Read
SHARE

Calibration of Oxygen Analyzer

Calibration of most types of oxygen analyzers can be performed manually or by an automatic calibration option. The analyzer controller can be programmed to perform the automatic calibration at specific time intervals, review the calibration results and if they are within programmable parameters, update the controller with the new calibration data.

Contents
Calibration of Oxygen AnalyzerTo CalibrateTroubleshooting of Oxygen AnalyzerSampling SystemSensor

The procedures are the same for manual or automatic calibrations with the exception that in manual mode, the calibration data has to be manually accepted after the calibration runs.

The calibration procedure is very similar to each type of analyzer, therefore, it is recommended to following the manufacturer’s procedure. The calibration procedures are typically programmer orientated and involve two calibrations gases with the results adjusting the slope of the calibration line within the software.

To Calibrate

  • Zero and span gases are connected to the analyzer.
  • With the zero-gas flowing through the analyzer, allow the readings to stabilize, then correct the reading to equal the zero gas concentration.
  • With the span-gas flowing through the analyzer, allow the readings to stabilize, then correct the reading to equal the span gas concentration.

Calibration and Troubleshooting of Oxygen Analyzer

Troubleshooting of Oxygen Analyzer

Faults with the oxygen analyzer output can be generated from several courses:

Sampling System

    • The sampling system should be periodically checked for proper flows and pressures and filtration and moisture removal.
    • Low readings are an indication of improper sampling.

Sensor

Zirconia Cell

The inaccurate readings could result from:

  • Heater temperature thermocouples
  • Heater itself
  • Heater temperature controller
  • Contaminated electrodes
  • Contaminated zirconia cell

All items should be checked according to the manufacturer’s procedures and diagnostics

Thermoparamagnetic

The inaccurate readings could result from:

  • Thermistors
  • Electronic sensing circuit
  • Auto-zero function
  • Changing magnetic field

All items should be checked according to the manufacturer’s procedures and diagnostics

Glass Sphere Paramagnetic

The inaccurate readings could result from:

  • Electronic sensing circuit
  • Light emitter
  • Photodiodes
  • Changing magnetic field

All items should be checked according to the manufacturer’s procedures and diagnostics

Electronics

All analyzers are equipped with extensive self-diagnostics and troubleshooting guides by the manufacturer, which should be followed for any alarms or inaccurate readings.

Author: Mohammed Khaleel

Read Next:

  • Heat Detector Testing
  • Fiber Optic Cable
  • H2S Gas Detector Calibration
  • Calibration of pH Analyzer
  • Purging of Instruments
Don't Miss Our Updates
Be the first to get exclusive content straight to your email.
We promise not to spam you. You can unsubscribe at any time.
Invalid email address
You've successfully subscribed !

Continue Reading

Bently Nevada Vibration Probes Functional Testing
Control Valve Stroke Test Procedure
What is calibration?
What is Reliability Centered Maintenance? (RCM)
Kiln Inlet Gas Analyzer
Load Cell Healthiness Checks
Share This Article
Facebook Whatsapp Whatsapp LinkedIn Copy Link
Share
2 Comments
  • Stephen kabwe says:
    June 16, 2022 at 8:06 pm

    Wonderful stuff here I love it

    Reply
  • B. PRAKASH says:
    March 16, 2023 at 10:08 pm

    Sir, I received concentrator for service, I found the concentrator pcb is defective. In that pcb memory ic is short. I could not do anything returned to the customer, without repair. The ic is available but I don’t software.

    Reply

Leave a Reply Cancel reply

Your email address will not be published. Required fields are marked *

Stay Connected

128.3kFollowersLike
69.1kFollowersFollow
210kSubscribersSubscribe
38kFollowersFollow

Categories

Explore More

Smart Transmitter Calibration Tutorial Part 3
Types of Calibration
Dissolved Oxygen Analyzer Interview Questions & Answers
Temperature Sensor calibration procedure
History of Measurement
Ultrasonic Level Transmitter Principle, Limitations, Calibration and configuration
Instrument Turndown Ratio
Displacer Level Transmitter Calibration using Water & Process Liquid

Keep Learning

Difference between Calibration and Ranging

Difference between Calibration and Ranging

Instrument Maintenance and Calibration System

Instrument Maintenance and Calibration System

conductivity transmitter

Conductivity Transmitter Calibration Procedure

SMART Transmitters Block Diagram

Smart Transmitters LRV & URV

pH Analyzer Calibration and Troubleshooting

Troubleshooting pH Analyzer Common Problems

Impulse Line Pressure Testing Procedure

Impulse Line Pressure Testing Procedure

ORP-sensor-calibration

Oxidation-Reduction Potential (ORP) Sensor Calibration Procedure

Selection of Weighing Balance

Selection of Weighing Balance and Weight for Calibration

Learn More

PLC Program to Count and Pack Parts from conveyor

Count and Pack Objects from Conveyor using PLC Ladder Logic

full-wave-bridge-rectifier-with-voltage-regulator

Basics of Voltage Regulators

Sampling Circuits Objective Questions

Sampling Circuits Objective Questions

Simple Pressure Repeater

Simple Pressure Repeater

What is Profibus

What is Profibus? – A Comprehensive Guide for Beginners

Relief Valve

Pressure Relief Valve Principle

Electric Motors Efficiency Classes

Electric Motors Efficiency Classes

Permanent magnet moving coil instruments (PMMC)

Permanent Magnet Moving Coil Instruments (PMMC)

Menu

  • About
  • Privacy Policy
  • Copyright

Quick Links

  • Learn PLC
  • Helping Hand
  • Part Time Job

YouTube Subscribe

Follow US
All rights reserved. Reproduction in whole or in part without written permission is prohibited.
Welcome Back!

Sign in to your account

Username or Email Address
Password

Lost your password?