CALIBRATION PROCEDURE FOR DIGITAL CONTROL VALVE
The below calibration procedure for Digital Control Valve. Make : Fisher, Model : DVC 6000
Also Read : Digital Control Valve Principle
Tools Required:
1. 375 Field communicator
2. Tools kit
3. DATA sheet
Calibration Procedure
TroubleShooting:
For fisher control valve adjust travel tuning set to lower alphabet ( If setting was D we have to select C).
For other valves we have adjust proportional gain
For fisher control valve adjust pressure tuning set to lower alphabet ( If setting was D we have to select C).
For other valves we have adjust proportional gain
NOTE:
Relay – A = Double Acting
Relay –B = Reverse acting
Relay –C = Forward & Reverse acting
The conveyor sorting machine is widely used in the packing industries using the PLC program…
Learn the example of flip-flop PLC program for lamps application using the ladder logic to…
In this article, you will learn the STAR DELTA programming using PLC controller to start…
Lube oil consoles of rotary equipment packages in industrial process plants are usually equipped with…
Rotating equipment packages such as pumps, compressors, turbines need the lube oil consoles for their…
This article explains how to blink lights in ladder logic with a detailed explanation video…