Calibration

Stability and Uniformity of Temperature Bath

Stability and uniformity are important characteristics of Temperature Bath.

Before buying any Temperature Bath, always remember to check for its Stability and Uniformity.

Stability of Temperature Bath

Temperature Stability measures how precisely an instrument maintains a set-point temperature.

Temperature Stability is expressed as the difference between the highest & lowest temperature values of the temperature sensors after achieving stability.

Procedure to find Stability of Temperature Bath

  • Placed the RTD sensor at the center of the temperature bath.
  • Set the required temperature of the temperature bath.
  • Let the bath stabilized at required temperature. ( Around 30 minutes after set point is attained)
  • Note down the readings of master RTD after every minute for 10 minutes.
  • The difference between the highest and lowest reading is the stability of the temperature bath.
Stability and Uniformity of Temperature Bath

Uniformity of Temperature Bath

Temperature Uniformity measures the potential variation in temperature between different points.

Temperature bath having a fan for circulation to have good uniformity.

Uniformity is of two types – Horizontal Uniformity and Vertical Uniformity.

Vertical Uniformity

Vertical Uniformity is calculated by placing the sensors at different immersion lengths of the temperature bath and 1 sensor placed in the center.

Horizontal Uniformity

Horizontal  Uniformity is calculated by placing the sensors at four corners of the bath at the same immersion length of temperature bath and 1 sensor placed in the center.

Procedure to find Uniformity of Temperature Bath

  • Place the RTD sensor.
  • Set the required temperature of the temperature bath.
  • Let the bath stabilized at required temperature. ( Around 30 minutes after set point is attained)
  • Note down the readings after every minute for 10 minutes for all five sensors for each points.
  • The difference between the center and other sensor’s maximum deviation will be taken as uniformity of temperature bath.

If you liked this article, then please subscribe to our YouTube Channel for Instrumentation, Electrical, PLC, and SCADA video tutorials.

You can also follow us on Facebook and Twitter to receive daily updates.

Read Next:

You've successfully subscribed !

View Comments

  • Thanks for sharing this amazing knowledge about Temperature Bath. Nice and engaging content.

Share

Recent Articles

  • PLC Tutorials

Conveyor Sorting Machine PLC Program with Calculation Function

The conveyor sorting machine is widely used in the packing industries using the PLC program…

5 days ago
  • PLC Tutorials

Example of Flip-Flop PLC Program for Lamps Application

Learn the example of flip-flop PLC program for lamps application using the ladder logic to…

5 days ago
  • PLC Tutorials

STAR DELTA Programming using PLC Controller

In this article, you will learn the STAR DELTA programming using PLC controller to start…

5 days ago
  • Instrumentation Design

Single and Three Phase Immersion Electrical Heaters Wiring Circuits

Lube oil consoles of rotary equipment packages in industrial process plants are usually equipped with…

2 days ago
  • Instrumentation Design

Immersion Heaters in Lube Oil Consoles

Rotating equipment packages such as pumps, compressors, turbines need the lube oil consoles for their…

4 days ago
  • PLC Tutorials

How to Blink Lights in Ladder Logic?

This article explains how to blink lights in ladder logic with a detailed explanation video…

4 weeks ago