Multiple Choice Questions

Special Designs of Ion Selective Electrodes Questions & Answers

Special Designs of Ion Selective Electrodes Questions & Answers

1. Which of the following gas permeable membrane is used for ammonia gas sensing electrode?

a) Silicon rubber
b) Microporous Teflon membrane
c) Fluorocarbon
d) Polythene

Answer: c

Explanation: Gas permeable membrane used for ammonia sensing electrode is fluorocarbon membrane. For each gas different membranes are used.

2. Which of the following gas permeable membrane is used for carbon dioxide gas sensing electrode?

a) Silicon rubber
b) PVC membrane
c) Fluorocarbon
d) Polythene

Answer: a

Explanation: Gas permeable membrane used for carbon dioxide sensing electrode is silicon rubber membrane. For each gas different membranes are used.

3. The response time of biocatalytic membrane electrode is poor when used with complex organic molecules.

a) True
b) False

Answer: b

Explanation: The response time of biocatalytic membrane electrode is good even when used with complex organic molecules. But, the membrane is costly.

4. Which of the following reference electrode is placed in the solution of carbon dioxide electrode?

a) Calomel electrode
b) Silver/silver chloride electrode
c) Mercury/mercury sulphate electrode
d) Glass electrode

Answer: b

Explanation: A silver/silver chloride reference electrode is placed in the solution. A glass electrode is also placed in the solution but it is not a reference electrode.

5. Biocatalytic membrane electrode cannot be used for the measurement of which of the following gases?

a) Ammonia
b) Carbon dioxide
c) Hydrogen
d) Nitrogen

Answer: d

Explanation: Biocatalytic membranes cannot be used for the measurement of nitrogen. It can be used for the measurement of other three gases.

6. The biocatalytic membrane used in ammonia selective electrode is which of the following?

a) Urea
b) Urease
c) Acrylamide
d) Polyacrylamide

Answer: b

Explanation: The biocatalytic membrane used in ammonia selective electrode is urease. It is used for the detection of ammonia.

7. Which of the following materials are used as gel to which biocatalytic membrane is fixed in ammonia selective electrode?

a) Urea
b) Urease
c) Acrylamide
d) Polyacrylamide

Answer: c

Explanation: The material used as gel to which biocatalytic membrane is fixed in ammonia selective electrode is acrylamide. This is attached to the ion selective electrode directly.

8. The biocatalytic membrane is attached to the glass electrode using which of the following materials?

a) Nylon mesh
b) Teflon
c) Silicon rubber
d) Polythene

Answer: a

Explanation: The biocatalytic membrane is attached to the glass electrode using nylon mesh. Cellophane film can also be used.

9. The response time of the biocatalytic electrode is a function of thickness of the enzyme layer.

a) True
b) False

Answer: a

Explanation: The response time of the biocatalytic electrode is a function of thickness of the enzyme layer. It is also a function of diffusion of the layer in the sample solution.

10. In carbon dioxide electrode, the membrane separates which of the following?

a) Sodium carbonate, magnesium chloride
b) Magnesium hydrogen carbonate, sodium chloride
c) Sodium hydrogen carbonate, sodium chloride
d) Magnesium carbonate, magnesium chloride

Answer: b

Explanation: In carbon dioxide electrode, the membrane separates sodium hydrogen carbonate and sodium chloride. Sodium hydrogen carbonate is the internal electrolyte.

You've successfully subscribed !
Share

Recent Articles

Electrical Drives – Modes, Types, Speed Control Applications
  • Electrical Basics

Electrical Drives – Modes, Types, Speed Control Applications

Electrical Drives control the motion of electric motors. Motion control is required in industrial and…

4 days ago
  • PLC Tutorials

PLC Ladder Logic Design: Control 3 Motors with Toggle Switch

PLC ladder logic design to control 3 motors with toggle switch and explain the program…

3 weeks ago
  • PLC Tutorials

VFD Simulator Download – Free Yaskawa V1000 Software

VFD simulator download: Master the online tool from the Yaskawa V1000 & programming software for…

4 days ago
  • PLC Tutorials

Conveyor Sorting Machine PLC Program with Calculation Function

The conveyor sorting machine is widely used in the packing industries using the PLC program…

7 days ago
  • PLC Tutorials

Example of Flip-Flop PLC Program for Lamps Application

Learn the example of flip-flop PLC program for lamps application using the ladder logic to…

1 week ago
  • PLC Tutorials

STAR DELTA Programming using PLC Controller

In this article, you will learn the STAR DELTA programming using PLC controller to start…

1 week ago