Multiple Choice Questions

Ion Spectroscopy Questions & Answers

Ion Spectroscopy Questions & Answers

1. In Ion spectroscopy, the primary ion is usually which of the following?

a) Inert gas ion
b) Halogen ion
c) Oxygen group gas ion
d) Hydrogen

Answer: a

Explanation: In Ion spectroscopy, the primary ion is usually an inert gas ion. When primary ion having particular kinetic energy is incident on a surface certain phenomena can occur.

2. The kinetic energy of the primary ion should be in which of the following range?

a) 0.1-4 keV
b) 0.2-1 keV
c) 0.4-2 keV
d) 0.3-5 keV

Answer: d

Explanation: The kinetic energy of the primary ion should be 0.3-5 keV. These primary ions are made incident on the surface.

3. If the primary ion is elastically scattered, the kinetic energy of the reflected primary ion will depend on which of the following?

a) Charge of the primary ion
b) Charge of the surface ion
c) Mass of the surface ion
d) Number of surface ions

Answer: c

Explanation: If the primary ion is elastically scattered, the kinetic energy of the reflected primary ion will depend on the mass of the surface ion. The reflected ion is measured by ISS.

4. Which of the following causes the phenomena of sputtering?

a) Primary ion gets embedded in the solid
b) Primary ion is elastically scattered
c) Primary ion is reflected
d) Primary ion is refracted

Answer: a

Explanation: Primary ion may penetrate through few layers of the surface and get embedded in the solid. This causes scattering.

5. The fragments formed during sputtering can be either neutral atoms or ions. The ions can only be positive.

a) True
b) False

Answer: b

Explanation: The fragments formed during sputtering can be either neutral atoms or ions. The ions can be positive or negative.

6. Which of the following is the energy range of ISS?

a) 1 keV
b) 2 keV
c) 4 keV
d) 8 keV

Answer: a

Explanation: The energy range of the ISS is 1 keV. ISS stands for ion scattering spectroscopy.

7. Which of the following is the spectral range of SIMS?

a) 0-10 amu
b) 0-100 amu
c) 0-500 amu
d) 0-1000 amu

Answer: c

Explanation: The spectral range of SIMS is 0-500 amu. SIMS stands for secondary ion mass spectroscopy.

8. Both ISS and SIMS have depth profiling capability.

a) True
b) False

Answer: a

Explanation: Both ISS and SIMS have depth profiling capability. ISS is slow in depth profiling but SIMS is rapid.

9. Which of the following is denotes the absolute quantitative analysis of SIMS?

a) 30%
b) 70%
c) 50%
d) Not possible

Answer: d

Explanation: Absolute quantitative analysis is not possible in SIMS. It is possible in ISS.

10. Which of the following is the amount of matrix effect that occurs in SIMS?

a) Very low
b) Low
c) Some
d) Severe

Answer: d

Explanation: Severe matrix effects occur in SIMS. Some matrix effects occur in ISS.

11. Which of the following is the x-y resolution of ISS?

a) 1 µ
b) 10 µ
c) 100 µ
d) 1000 µ

Answer: c

Explanation: The x-y resolution of ISS is 100 µ. It has poor x-y resolution.

12. Which of the following is the x-y resolution of SIMS?

a) 1 µ
b) 10 µ
c) 100 µ
d) 1000 µ

Answer: a

Explanation: The x-y resolution of SIMS is 1 µ. It has this resolution when used with ion microprobe.

You've successfully subscribed !
Share

Recent Articles

PLC Ladder Logic Design: Control 3 Motors with Toggle Switch
  • PLC Tutorials

PLC Ladder Logic Design: Control 3 Motors with Toggle Switch

PLC ladder logic design to control 3 motors with toggle switch and explain the program…

3 weeks ago
  • PLC Tutorials

VFD Simulator Download – Free Yaskawa V1000 Software

VFD simulator download: Master the online tool from the Yaskawa V1000 & programming software for…

4 days ago
  • PLC Tutorials

Conveyor Sorting Machine PLC Program with Calculation Function

The conveyor sorting machine is widely used in the packing industries using the PLC program…

7 days ago
  • PLC Tutorials

Example of Flip-Flop PLC Program for Lamps Application

Learn the example of flip-flop PLC program for lamps application using the ladder logic to…

7 days ago
  • PLC Tutorials

STAR DELTA Programming using PLC Controller

In this article, you will learn the STAR DELTA programming using PLC controller to start…

1 week ago
  • Instrumentation Design

Single and Three Phase Immersion Electrical Heaters Wiring Circuits

Lube oil consoles of rotary equipment packages in industrial process plants are usually equipped with…

4 days ago