Reading: Calibration of Weight using the ABBA Method
Share
Notification Show More
Font ResizerAa
Inst ToolsInst Tools
Font ResizerAa
  • Courses
  • Design
  • PLC
  • Interview
  • Control System
Search
  • Courses
  • Automation
    • PLC
    • Control System
    • Safety System
    • Communication
    • Fire & Gas System
  • Instrumentation
    • Design
    • Pressure
    • Temperature
    • Flow
    • Level
    • Vibration
    • Analyzer
    • Control Valve
    • Switch
    • Calibration
    • Erection & Commissioning
  • Interview
    • Instrumentation
    • Electrical
    • Electronics
    • Practical
  • Q&A
    • Instrumentation
    • Control System
    • Electrical
    • Electronics
    • Analog Electronics
    • Digital Electronics
    • Power Electronics
    • Microprocessor
Follow US
All rights reserved. Reproduction in whole or in part without written permission is prohibited.
Inst Tools > Blog > Calibration > Calibration of Weight using the ABBA Method

Calibration of Weight using the ABBA Method

In this article, you will learn the calibration procedure of weight using the ABBA method.

Last updated: September 28, 2022 3:24 pm
Saurabh Vaidya
Calibration
1 Comment
Share
2 Min Read
SHARE

In this article, you will learn the calibration procedure of weight using the ABBA method.

Calibration of Weight

Where A is Standard Weight and B is UUC weight (UUC is Unit Under Calibration)

Select an appropriate standard weight for calibration of UUC Weights.

i.e one class higher standard weight is required for UUC weight Calibration.

For F1 class calibration of weights, E2 class weight is required as standard weight

Calibration of Weight using the ABBA Method

Weight Calibration using ABBA
  1. Keep Standard weight A on weighing Balance and note down the reading.
  2. Keep UUC Weight B on weighing Balance and note down the reading. Remove weight from the weighing balance.
  3. Again, Keep UUC Weight B on weighing Balance and note down the reading.
  4.  Keep Standard weight A on weighing Balance and note down the reading.

This completes one set of reading.

ABA Method and AB1…..BnA are similar Methods.

ABBA          Standard   ______ UUC  ______ UUC  ______ Standard

ABA             Standard ______  UUC  ______ Standard

AB1…..BnA   Standard  ______  UUC ______ UUC  ______  UUC (n times) ______ Standard

Calibration of Weight using the ABBA Method

The equation for Determination of Conventional Mass

mct = mcr(1+C)+Δmw

Where

mct is the Conventional mass of test weight

mcr is the Conventional mass of reference weight

C is the Buoyancy correction factor

Δmw is the difference in weight by the ABBA method

For the same material weight with the same density, the Buoyancy correction factor is Zero.

ΔM is calculated as follows for ABBA Method

Let’s consider ABBA as A1B1B2A2

ΔM = [(B1-A1) +(B2-A2)] /2

A1 = first reading of reference weight

B1 = first reading of test weight

B2 = second reading of test weight

A2 = second reading of reference weight

Don't Miss Our Updates
Be the first to get exclusive content straight to your email.
We promise not to spam you. You can unsubscribe at any time.
Invalid email address
You've successfully subscribed !

Continue Reading

How to Calibrate pH Electrode?
Step by Step Guide for Gas Flow Meter Calibration
Oxidation-Reduction Potential (ORP) Sensor Calibration Procedure
Smart Transmitters LRV & URV
What is a Dry Block? – Parts of the Temperature Calibrator
Calibration of Temperature Sensor with Indicator
Share This Article
Facebook Whatsapp Whatsapp LinkedIn Copy Link
Share
1 Comment
  • Amaresh Senthilkumar says:
    October 4, 2023 at 6:05 pm

    Your articles are very useful and especially the uncertainty of the pressure gauge is awesome. you are explaining clearly step by step.. is there any possibility of uploading the linear instruments uncertainty budget kindly upload it.

    Reply

Leave a Reply Cancel reply

Your email address will not be published. Required fields are marked *

Stay Connected

128.3kFollowersLike
69.1kFollowersFollow
210kSubscribersSubscribe
38kFollowersFollow

Categories

Explore More

Single Position Calibration of Temperature Bath
H & B Gas Analyzer Principle and Calibration Procedure
How to Calibrate Load Cells in Weighing Applications?
Conductivity Transmitter Calibration Procedure
Difference between Accuracy and Precision
Displacer Level Transmitter Calibration using Water & Process Liquid
Bently Nevada Vibration Probes Functional Testing
RTD Calibration

Keep Learning

Calibrate Eddy Current Conductivity Instrument

How to Calibrate the Eddy Current Conductivity Meter?

NIST Traceability

NIST Traceability

Multimeter Digits

Multimeter – What Do Digits and Counts Mean?

Impulse Line Pressure Testing Procedure

Impulse Line Pressure Testing Procedure

Differential Pressure Transmitter Manifold

Field Instrument Calibration Errors

Why Calibration is Important

Why Calibration is Important?

Pig Signalling Switches Functional Testing

Pig Signalling Switches Functional Testing

Temperature Switch Calibration

How to Calibrate Temperature Switch? – Instrument Basics

Learn More

Fluorescent Screen of CRT

CRT Fluorescent Screen

Cascade-Temperature-Control-System

Cascade Temperature Control System

PLC Program for Solenoid, Pilot Lamp, and Switch with Alarm

PLC Program for Solenoid, Pilot Lamp, and Switch with Alarm

Full Wave Bridge Rectifier Operation

Full Wave Bridge Rectifier Operation

PLC Commissioning and Testing

PLC Commissioning and Testing Procedure (Programmable Logic Controller)

Safety Valve

Compare Relief Valve and Safety Valve

baffle nozzle Operation

Practical Pneumatic Instruments

Power Electronics Objective Questions

Thyristor Mounting Objective Questions and Answers

Menu

  • About
  • Privacy Policy
  • Copyright

Quick Links

  • Learn PLC
  • Helping Hand
  • Part Time Job

YouTube Subscribe

Follow US
All rights reserved. Reproduction in whole or in part without written permission is prohibited.
Welcome Back!

Sign in to your account

Username or Email Address
Password

Lost your password?