Inst ToolsInst ToolsInst Tools
  • Courses
  • Videos
  • Q & A
    • Interview
      • Instrumentation
      • Electronics
      • Electrical
      • Practical Questions
    • MCQ
      • Instrumentation MCQ
      • Electrical MCQ
      • Electronics MCQ
      • Control Systems MCQ
      • Analog Electronics MCQ
      • Digital Electronics MCQ
      • Power Electronics MCQ
      • Microprocessor MCQ
      • Multiple Choice Questions
  • EE
    • Electronics
      • Electronics Q & A
      • Electronic Basics
      • Electronic Devices & Circuits
      • Electronics Animation
      • Digital Electronics
    • Electrical
      • Electrical Basics
      • Electrical Q & A
      • Power Electronics
      • Electrical Machines
      • Electrical Animation
      • Power Systems
      • Switchgear & Protection
      • Transmission & Distribution
  • Measure
    • Control Valves
    • Calibration
    • Temperature
    • Pressure
    • Flow
    • Level
    • Analyzers
    • Switches
    • Vibration
    • Solenoid Valve
  • Control
    • PLC Tutorials
    • Control Systems
    • Safety Instrumented System (SIS)
    • Communication
    • Fire & Gas System
  • More
    • Design
    • Tools
    • Animation
    • Basics
    • Formulas
    • Standards
    • TextBooks
    • Common
    • Software
    • Excel Tools
    • Erection & Commissioning
    • Process Fundamentals
    • Videos
    • Books
Search
All rights reserved. Reproduction in whole or in part without written permission is prohibited.
Reading: Capillary Tubing Pressure Sensor Temperature & Elevation Problems
Share
Notification Show More
Font ResizerAa
Inst ToolsInst Tools
Font ResizerAa
  • Courses
  • PLC Tutorials
  • Control Systems
Search
  • Courses
  • Videos
  • Q & A
    • Interview
    • MCQ
  • EE
    • Electronics
    • Electrical
  • Measure
    • Control Valves
    • Calibration
    • Temperature
    • Pressure
    • Flow
    • Level
    • Analyzers
    • Switches
    • Vibration
    • Solenoid Valve
  • Control
    • PLC Tutorials
    • Control Systems
    • Safety Instrumented System (SIS)
    • Communication
    • Fire & Gas System
  • More
    • Design
    • Tools
    • Animation
    • Basics
    • Formulas
    • Standards
    • TextBooks
    • Common
    • Software
    • Excel Tools
    • Erection & Commissioning
    • Process Fundamentals
    • Videos
    • Books
Follow US
All rights reserved. Reproduction in whole or in part without written permission is prohibited.
Inst Tools > Blog > Pressure Measurement > Capillary Tubing Pressure Sensor Temperature & Elevation Problems

Capillary Tubing Pressure Sensor Temperature & Elevation Problems

Last updated: July 14, 2019 5:07 pm
Editorial Staff
Pressure Measurement
No Comments
Share
4 Min Read
SHARE

Capillary tubing provides a convenient method to separate a pressure instrument such as a gauge or transmitter from the actual process connection, allowing the instrument to be installed in a more accessible location.

Contents
1. Temperature effects2. Elevation effects3. Response time

However, users should be aware of the effects that capillary tubing has on the operation of the pressure sensing element.

Capillary tubing introduces three effects on the performance of pressure-sensing instruments:

1. Temperature effects

Temperature changes cause the liquid inside the capillary tube to expand and contract, changing the volume of the fill fluid.

The resulting error in the pressure reading is a function of the total volume of the tubing, pressure instrument, and isolator ring.

Because the rubber sleeve in the isolator ring has a much lower modulus of elasticity compared to a diaphragm seal, it can absorb most of the volumetric change resulting from temperature differences throughout the usable temperature range for isolator rings.

A typical error in gauge reading through a temperature swing from 0°F to 120°F is about ½ psi depending on isolator ring size and gauge type.

This is roughly a quarter the error expected with a standard 60 mm stainless steel diaphragm seal

2. Elevation effects

As you change the elevation of the gauge with respect to the isolator ring, you introduce an elevation error. This error is due to the static pressure of the liquid in the capillary tube.

The change in gauge reading caused by elevation changes of the pressure sensing instrument can be calculated in advance using the following equation:

Capillary Tubing Pressure Equation

Our standard fill fluid is a silicone oil with a specific gravity = 0.967 at 77°F.

Observe polarity: If the gauge is above the isolator ring, then the elevation term in the above equation is positive; if the gauge is below the isolator ring, the elevation is negative.

If the gauge or transmitter has a zero adjust capability, the elevation error can be eliminated completely by re-setting the zero adjustment to compensate for the elevation change.

Effects of Capillary Tubing on Pressure Measurement

3. Response time

Capillary tubing introduces a response time lag in the instrument reading. This delayed reaction time is influenced by:

  • Length of the capillary tube
  • Internal diameter of the capillary tube
  • Control volume of the pressure-sensing instrument
  • Viscosity of the fill fluid, including temperature effects fill fluid viscosity

The control volume of a pressure-sensing instrument such as a gauge or transmitter is defined as the change in volume required to deflect the bourdon tube or sensing diaphragm from zero to 100% reading.

The smaller the control volume the better the performance. Instruments with smaller control volumes exhibit less temperature error and time lag than instruments with greater volume.

As a general rule, higher range instruments have a smaller control volume than lower range instruments; for example a 100-psi gauge has a much smaller control volume than a 15-psi gauge.

Another general rule is that electronically amplified devices have a smaller control volume than mechanical devices; again, to use an example, an electronic transmitter has about 1/100th the control volume of a bourdon tube gauge of the same range.

Capillary Tubing Pressure Sensor Temperature & Elevation Problems

Response time is defined as the time, in seconds, for the pressure-sensing instrument to register 63.2% of a step change in pressure.

The reason the definition is not based on a 100% change is because, theoretically, it takes an infinite amount of time for any pressure-sensing instrument to respond to a step change in pressure.

That’s because the gauge pointer moves slower and slower as it gets closer to the actual pressure.

Author : David Gardellin, P. Eng, Onyx Valve Co

Don't Miss Our Updates
Be the first to get exclusive content straight to your email.
We promise not to spam you. You can unsubscribe at any time.
Invalid email address
You've successfully subscribed !

Recommended Articles

Pressure Gauge Accessories
Pressure Gauge Problems
Pressure Transmitter Troubleshooting Tips
Why are Impulse Lines often Installed in a Sloping Manner?
Pressure Transmitter Circuit
Problem on Water Filter Discharge Pressure Transmitter and Gauge
Share This Article
Facebook Whatsapp Whatsapp LinkedIn Copy Link
Share
Leave a Comment

Leave a Reply Cancel reply

Your email address will not be published. Required fields are marked *

Stay Connected

128.3kFollowersLike
69.1kFollowersFollow
208kSubscribersSubscribe
38kFollowersFollow

Categories

Recent Comments

  • William Snyder on Top Non-PLC Certification Courses for Automation Professionals
  • Kamli on Top Free PLC Software
  • Guifty Shimica on Top Non-PLC Certification Courses for Automation Professionals
  • Guifty Shimica on Top Non-PLC Certification Courses for Automation Professionals

Related Articles

Pressure Transmitter Working Principle

Gage and Absolute Pressure Transmitters Principle

Differential Pressure Switch Principle

Differential Pressure Switch Principle

U-tube Manometer unEqual Limb Type

U-tube Manometer with equal & unequal limb types

Instrument Protection Methods: Seals

Pressure Transmitters Filled impulse lines

Pressure Transmitters Filled impulse lines

Gauge Display Scale

How to Read Gauge Analog Scales

Closed Tank Liquid Level Measurement using Pressure Transmitter

Inferential Measurement based on Pressure

Pressure Detection Circuit

Pressure Detection Circuit

More Articles

GATE Instrumentation Questions and Answers

GATE Instrumentation Questions and Answers

Catalytic Gas Sensor Working

Catalytic Type Gas Sensor Principle

Ultrasonic Sensors Objective Questions

Ultrasonic Sensors Objective Questions

Hybrid Solar Power System

Difference Between Normal Inverter and Solar Inverter

In-Situ Zirconia Oxygen Analyzer

Oxygen (O2) Measurement in Flue Gas

Level Gauges

Level Gauge Design

Pneumatic Counter for Counting the Strokes of a Cylinder

Pneumatic Counter for Counting the Strokes of a Cylinder

OPC Server

What is the OPC Server?

Follow US
All rights reserved. Reproduction in whole or in part without written permission is prohibited.
Welcome Back!

Sign in to your account

Username or Email Address
Password

Lost your password?